Sign In | Join Free | My burrillandco.com
Home > ELISA Test Kit >

OXTR Sandwich Enzyme Linked Immunosorbent Assay Test Kit For Research

Shanghai Korain Biotech Co., Ltd
Contact Now
    Buy cheap OXTR Sandwich Enzyme Linked Immunosorbent Assay Test Kit For Research from wholesalers
     
    Buy cheap OXTR Sandwich Enzyme Linked Immunosorbent Assay Test Kit For Research from wholesalers
    • Buy cheap OXTR Sandwich Enzyme Linked Immunosorbent Assay Test Kit For Research from wholesalers

    OXTR Sandwich Enzyme Linked Immunosorbent Assay Test Kit For Research

    Ask Lasest Price
    Brand Name : BT Lab
    Model Number : Cat.No E0291Ca
    Certification : CE, ISO9001:2005, MSDS
    Price : Negotiation
    Supply Ability : Western Union, T/T
    Delivery Time : 1-3 business days, bulk order within one week
    • Product Details
    • Company Profile

    OXTR Sandwich Enzyme Linked Immunosorbent Assay Test Kit For Research

    Canine Strong Sensitivity and Specificity OXTR ELISA Test Kit With Oem Service 96 Well Plate


    Cat.No E0291Ca


    Precision

    Intra-Assay Precision (Precision within an assay) Three samples of known concentration were tested on one plate to assess intra-assay precision.

    Inter-Assay Precision (Precision between assays) Three samples of known concentration were tested in separate assays to assess inter-assay precision.

    CV(%) = SD/mean x 100

    Intra-Assay: CV<8%

    Inter-Assay: CV<10%


    Standard Curve Range: 0.2ng/ml - 60ng/ml

    Sensitivity: 0.097ng/ml

    Size: 96 wells

    Storage: Store the reagents at 2-8°C. For over 6-month storage refer to the expiration date keep it at -20°C. Avoid repeated thaw cycles. If individual reagents are opened it is recommended that the kit be used within 1 month.

    *This product is for research use only, not for use in diagnosis procedures. It’s highly recommend to read this instruction entirely before use.


    Assay Principle

    This ELISA Test Kit is an Enzyme-Linked Immunosorbent Assay (ELISA). The plate has been pre-coated with Canine OXTR antibody. OXTR present in the sample is added and binds to antibodies coated on the wells. And then biotinylated Canine OXTR Antibody is added and binds to OXTR in the sample. Then Streptavidin-HRP is added and binds to the Biotinylated OXTR antibody. After incubation unbound Streptavidin-HRP is washed away during a washing step. Substrate solution is then added and color develops in proportion to the amount of Canine OXTR. The reaction is terminated by addition of acidic stop solution and absorbance is measured at 450 nm.


    Reagent Preparation

    All reagents should be brought to room temperature before use

    Standard Reconstitute the 120μl of the standard (64ng/ml) with 120μl of standard diluent to generate a 32ng/ml standard stock solution. Allow the standard to sit for 15 mins with gentle agitation prior to making dilutions. Prepare duplicate standard points by serially diluting the standard stock solution (32ng/ml) 1:2 with standard diluent to produce 16ng/ml, 8ng/ml, 4ng/ml and 2ng/ml solutions. Standard diluent serves as the zero standard(0 ng/ml). Any remaining solution should be frozen at -20°C and used within one month. Dilution of standard solutions suggested are as follows:


    32ng/mlStandard No.5120μl Original Standard + 120μl Standard Diluent
    16ng/mlStandard No.4120μl Standard No.5 + 120μl Standard Diluent
    8ng/mlStandard No.3120μl Standard No.4 + 120μl Standard Diluent
    4ng/mlStandard No.2120μl Standard No.3 + 120μl Standard Diluent
    2ng/mlStandard No.1120μl Standard No.2 + 120μl Standard Diluent

    Standard ConcentrationStandard No.5Standard No.4Standard No.3Standard No.2Standard No.1
    64ng/ml32ng/ml16ng/ml8ng/ml4ng/ml2ng/ml

    Wash Buffer Dilute 20ml of Wash Buffer Concentrate 25x into deionized or distilled water to yield 500 ml of 1x Wash Buffer. If crystals have formed in the concentrate, mix gently until the crystals have completely dissolved.


    Assay Procedure

    1. Prepare all reagents, standard solutions and samples as instructed. Bring all reagents to room temperature before use. The assay is performed at room temperature.

    2. Determine the number of strips required for the assay. Insert the strips in the frames for use. The unused strips should be stored at 2-8°C.

    3. Add 50μl standard to standard well. Note: Dont add antibody to standard well because the standard solution contains biotinylated antibody.

    4. Add 40μl sample to sample wells and then add 10μl anti-OXTR antibody to sample wells, then add 50μl streptavidin-HRP to sample wells and standard wells ( Not blank control well ). Mix well. Cover the plate with a sealer. Incubate 60 minutes at 37°C.

    5. Remove the sealer and wash the plate 5 times with wash buffer. Soak wells with at least 0.35 ml wash buffer for 30 seconds to 1 minute for each wash. For automated washing, aspirate all wells and wash 5 times with wash buffer, overfilling wells with wash buffer. Blot the plate onto paper towels or other absorbent material.

    6. Add 50μl substrate solution A to each well and then add 50μl substrate solution B to each well. Incubate plate covered with a new sealer for 10 minutes at 37°C in the dark.

    7. Add 50μl Stop Solution to each well, the blue color will change into yellow immediately.

    8. Determine the optical density (OD value) of each well immediately using a microplate reader set to 450 nm within 10 minuets after adding the stop solution.


    Calculation of Result

    Construct a standard curve by plotting the average OD for each standard on the vertical (Y) axis against the concentration on the horizontal (X) axis and draw a best fit curve through the points on the graph. These calculations can be best performed with computer-based curve-fitting software and the best fit line can be determined by regression analysis.